The IONIVAC ITR 90 and ITR 200 are broad-range vacuum gauges that combine a Bayard-Alpert hot cathode ionisation sensor with a Pirani sensor in a single compact transmitter. They provide continuous pressure measurement from 5 × 10⁻¹⁰ mbar up to atmospheric pressure (1000 mbar), making them ideal for analytical instruments, coating systems, vacuum furnaces, research applications, and general high-vacuum processes. The ITR 90 features a single hot-cathode filament, while the ITR 200 incorporates dual filaments for extended service life and increased operational reliability through automatic filament changeover.

 

They must not be used for measuring flammable or combustible gases in mixtures containing oxidants (e.g. atmospheric oxygen) within the explosion range.

ITR 200

The ITR 200 S is an optimized dual filament combination transmitter on the basis of the well-proven ITR 90. The combination of a hot cathode ionization sensor according to Bayard-Alpert and a Pirani sensor allows vacuum pressure measurements of non-ignitable gases and gas mixtures in the pressure range from 5 x 10-10 to 1000 mbar.

Upon request, the pressure can be displayed on an integrated display.

 

ITR 200 S (without display, 1 switching function) / ITR 200 S (with display, 1 switching function) / ITR 200 SP (with Profibus interface, 2 switching functions) / ITR 200 SD (with DeviceNet interface, 2 switching functions)

Features and Benefits

  • Service life increase and increased operational reliability through integration of a second hot cathode
  • Full coverage of the pressure range from 5 x 10-10 mbar to atmospheric pressure
  • High repeatability within the typical process pressure range of 10-2 to 10-8 mbar
  • Controlled switching on and switching off through the integrated dual Pirani optimized the service life of the yttrium-coated iridium cathodes
  • Compact design
  • Enclosed, stable electrode geometry in rugged metal casing
  • Efficient degassing through electron bombardment
  • Simple to install
  • ITR 200 S version with built-in display allows for stand-alone operation without the necessity for additional displays
  • RS 232 C interface

치수 높이 0.067
치수 길이 0.153
치수 너비 0.058
치수 코멘트 The dimensions length, width and height try to describe the outer shape of this particular product in its typical orientation when standing on the ground. We always recommend checking the detailed dimensions given in 2D drawings in the instruction manual or the 3D shapes given in 3D models.

압력 측정 최소값 5.000 × 10⁻⁸
최대 압력 측정 1.00

Level of Gauge Accuracy Medium
Classification Guage for Corrosive Gases No

주변 작동 온도 최대값 50.00
최대 주변 보관 온도 70.00
주변 작동 온도 최소 0.000
최소 주변 보관 온도 -20.00

최대 베이킹 온도 150.00