CERAVAC CTR 101 - DN16CF 1Torr

Part Number: 230328V01
i Replaces product 230328V02, 230328

Contact us for availability

Click For Availability

CERAVAC CTR

The CERAVAC CTR transmitters offer excellent accuracy and reproducibility. The CTR 100 and CTR 101 allow gas type independent pressure measurements and are able to tolerate bursts of pressure without suffering physical damage or calibration shifts. The robust sensor is suited for the most corrosive processes as the sensor is highly resistant to corrosion from common process chemicals. The sensor of the CTR 101 is internally heated and regulated to 45 °C to offer full-scale pressure ranges from 1000 to 0.1 Torr.

 

The CERAVAC CTR line is drop in replaceable for the CERAVAC CTR-N line with negligible impact.

Features and Benefits

  • Leading levels of accuracy (0.2%) achieved via the ceramic capacitive sensor
  • Excellent temperature compensation regardless of ambient conditions
  • Fast and accurate response times to pressure changes
  • Long-term stability: no calibration shifts after bursts of pressure
  • Measurement/display range: from 1000 Torr to 10-5, depending on the model

General Product Information

Cable Connector Comms Interface DSUB 15-PIN - Male
Connector Electrical Power DSUB 15-PIN - Male
Connection Inlet Flange DN 16 CF (1.33")
Dimension Height 0.082
Dimension Length 0.146
Dimension Width 0.082
IP Rating IP30
Power Supply Frequency DC Voltage
Power Supply Voltage DC 14-30V
Recommended Controller Graphix
Weight 0.897
Dimension Comments The dimensions length, width and height try to describe the outer shape of this particular product in its typical orientation when standing on the ground. We always recommend checking the detailed dimensions given in 2D drawings in the instruction manual or the 3D shapes given in 3D models.

Classification

Classification Product Pressure Range Medium Vacuum: 1 mbar - 1·10⁻³ mbar
Classification Principle of Measurement Capacitance diaphragm vacuum gauges
Level of Gauge Accuracy High
Classification Guage for Corrosive Gases Yes
Competitor Gauge Replacement Leybold - internal standard

Performance Data

Measurement Uncertainty 0.15% ± temperature effect
Power Consumption Max 12.00
Power consumption Max String while being heated ≤12 W; at operating temperature ≤8 W
Pressure Measurement Min 0.010
Pressure Measurement Max 133.00

Environmental Conditions

Temperature Ambient Operation Max 10.00
Temperature Ambient Storage Max 65.00
Temperature Ambient Operation Min 40.00
Temperature Ambient Storage Min -40.00

Material of Construction

Material Exposed to Vacuum Aluminium oxide (Al2O3) Stainless Steel AISI 316L

Connectivity - Communication Interface

Communication Interface Protocol CTR 100/101: 0-10V linear - -5 to +10.24 V
Gauge RS232 digital protocol
Communication Interfaces Customer 0-10 V (analog input - linear gauge)
Connection Electrical Comms Interface Connector DSUB 15-PIN - Male
Setpoint Quantity 2
Status Indicator LED

Variants

Certificated Pressure Measurement Device Standard OEM
Variant Gauges CTR 101 (heated version of CTR 100)
Type of Gauge Cable Type C (15-pin Sub-D to 15-pin Sub-D)

Operation Conditions - Limitations

Altitude Max 2000.00
Temperature Baking Max 110.00